Calculations and Patient Reports
14.7.4.7 Ao/LA
There are two methods to calculate Ao/LA (Aorta/Left Atrium) in the M mode.
One method is to calculate all items at a time and the other is to calculate one by one
.
•
To measure all Items at a Time
(chapter 14.7.4.7.1)
•
To measure the Items One by One
(chapter 14.7.4.7.2)
14.7.4.7.1
To measure all Items at a Time
1. After obtaining a feasible image, press the
[
Calc
]
key.
The Cardiac calculations menu appears in the menu area.
2. Select
[
Ao/LA
]
and select
[
All
]
.
Note:
The menu area shows which measurement should be performed.
(illuminated item e.g., Ao Root Dia)
3. Move + cursor to the Anterior Aortic Wall and press the right or
left trackball key
[
Set
]
.
4. Move + cursor to the Posterior Aortic Wall and press
[
Set
]
.
(Ao Root Diameter measurement)
5. Move + cursor to the Coronary Cusp and press
[
Set
]
to fix the mark.
6. Move + cursor to the Non Coronary Cusp and press
[
Set
]
. (Ao Cusp Separation measurement)
7. Move + cursor to the Posterior Aortic Wall and press
[
Set
]
to fix the mark.
8. Move + cursor to the Left Artrial Wall and press
[
Set
]
. (LA Diameter measurement)
14.7.4.7.2
To measure the Items One by One
To measure Ao Cusp Separation, Ao Root Diameter, LA Diameter one by one:
1. Press
the
[
Calc
]
key on the control panel.
The Cardiac calculations menu appears in the menu area.
2. Select
[
Ao/LA
]
.
3. Select the appropriate item among
[
Ao Root Dia
]
,
[
AoV Cusp Sep
]
or
[
LA Dia
]
.
4. Perform the measurements by using the trackball and the right or left trackball key
[
Set
]
.
Voluson
®
730Pro - Basic User Manual
105831 Rev. 0
14-33
Содержание Voluson 730Pro
Страница 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Страница 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Страница 22: ...CONTENTS This page intentionally left blank Voluson 730Pro Basic User Manual 18 CONTENTS 105831 Rev 0 ...
Страница 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Страница 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Страница 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Страница 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Страница 262: ...Volume Mode This page intentionally left blank Voluson 730Pro Basic User Manual 11 78 105831 Rev 0 ...
Страница 290: ...Basic Measurements This page intentionally left blank Voluson 730Pro Basic User Manual 13 16 105831 Rev 0 ...
Страница 411: ...Biopsy Setup 19 Biopsy Setup 19 2 19 1 To Program a Biopsy Line 19 3 Voluson 730Pro Basic User Manual 105831 Rev 0 19 1 ...
Страница 414: ...Biopsy Setup This page intentionally left blank Voluson 730Pro Basic User Manual 19 4 105831 Rev 0 ...
Страница 442: ...Probes and Biopsy This page intentionally left blank Voluson 730Pro Basic User Manual 20 28 105831 Rev 0 ...
Страница 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...
Страница 458: ...Connections 21 2 11 ECG preamplifier MAN Connection Voluson 730Pro Basic User Manual 21 16 105831 Rev 0 ...