PD Mode
9.3.8 Dynamic
Set
“Dynamic Set” controls the axial resolution of color in the display.
It adjusts the axial sample depth of color pixels.
When rotating counter-clockwise the color information is sharper (edges more pixely).
When rotation clockwise the color information is less sharp (edges smoothed).
Rotate the
[
Dyn.Set
]
digipot and select the desired dynamic range. Set 1 to Set 8.
1-4 color samples in axial direction shorter
4-8 color samples in axial direction larger
Remark:
Changes are only possible in write mode!
9.3.9 Smoothing
A temporal average is determined from several color images, so that different filter periods can be
selected for rising velocity and falling velocity.
Rotate the
[
Smooth
]
digipots and select the rise and fall filter.
RISE:
Filtering the rise velocity leads to noise suppression.
To be used with small laminar flows. Avoid quick movements of the probe, because the flow
is “built up” slowly. When displaying pulses the Rise-Filter must be set low.
FALL: This filter leads to prolongation of the displayed flow.
To be used with quick pulses (short "color flashes") to prolong them for better evaluation on
the monitor.
Remark:
Changes are only possible in write mode!
9.3.10
Exit the PD Submenu
Press the
[
Exit
]
key on the control panel to exit the PD Submenu.
Voluson® 730Pro - Basic User Manual
9-12
105831 Rev. 0
Содержание Voluson 730Pro
Страница 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Страница 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Страница 22: ...CONTENTS This page intentionally left blank Voluson 730Pro Basic User Manual 18 CONTENTS 105831 Rev 0 ...
Страница 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Страница 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Страница 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Страница 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Страница 262: ...Volume Mode This page intentionally left blank Voluson 730Pro Basic User Manual 11 78 105831 Rev 0 ...
Страница 290: ...Basic Measurements This page intentionally left blank Voluson 730Pro Basic User Manual 13 16 105831 Rev 0 ...
Страница 411: ...Biopsy Setup 19 Biopsy Setup 19 2 19 1 To Program a Biopsy Line 19 3 Voluson 730Pro Basic User Manual 105831 Rev 0 19 1 ...
Страница 414: ...Biopsy Setup This page intentionally left blank Voluson 730Pro Basic User Manual 19 4 105831 Rev 0 ...
Страница 442: ...Probes and Biopsy This page intentionally left blank Voluson 730Pro Basic User Manual 20 28 105831 Rev 0 ...
Страница 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...
Страница 458: ...Connections 21 2 11 ECG preamplifier MAN Connection Voluson 730Pro Basic User Manual 21 16 105831 Rev 0 ...