Basic Measurements
13.4.5
3D MultiPlane Measurement
This measuring program allows for volume determination of any organ which was stored by a volume
scan. Several parallel planes are laid through the organ and the areas of these planes are determined.
A measuring program calculates the volume from the measured areas and the distance between the
areas. The larger the number of areas, the more exact the volume calculation result will be.
Condition: A stored volume scan (Sectional Planes view).
To measure volume using the MultiPlane method.
1. Select the reference image in which the measurement is to be performed. (A, B or C)
2. Select
the
[
MultiPlane
]
item. The menu area changes to the 3D MultiPlane menu.
3. Select the first section through the body by rotating the
[
Ref. Slice
]
digipot
(make parallel sections through the reference image).
Note:
The first section should be set at the edge of the measured object.
4. Measure the area (proceed as for area measurement).
Position the start dot of the area to be surrounded with the trackball and store it. Surround the area
with the trackball, then press the right or left trackball key
[
Set
]
.
The area is calculated and displayed. The area may even be "zero" (dot at the edge).
5. Touch the key
[
Set
]
twice
!
6. Select the next parallel section with the
[
Ref. Slice
]
digipot and measure the area.
7. Repeat 5. and 6. until the edge of the measured object is reached.
Remarks:
•
The contour of the measured area is not erased if a new section is adjusted. From the deviation in
the new section it is possible to decide whether a new area should be marked.
With new marking the old contour is erased.
•
To call back the measured areas touch the
[
Prev / Next
]
flip switch.
•
The different sections can be chosen liberally, it is not necessary to follow a certain order.
•
The volume measurement is only possible in 3D mode or full-image mode (not in aspect mode).
•
To erase results, select the
[
Init
]
item.
Voluson
®
730Pro - Basic User Manual
13-10
105831 Rev. 0
Содержание Voluson 730Pro
Страница 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Страница 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Страница 22: ...CONTENTS This page intentionally left blank Voluson 730Pro Basic User Manual 18 CONTENTS 105831 Rev 0 ...
Страница 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Страница 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Страница 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Страница 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Страница 262: ...Volume Mode This page intentionally left blank Voluson 730Pro Basic User Manual 11 78 105831 Rev 0 ...
Страница 290: ...Basic Measurements This page intentionally left blank Voluson 730Pro Basic User Manual 13 16 105831 Rev 0 ...
Страница 411: ...Biopsy Setup 19 Biopsy Setup 19 2 19 1 To Program a Biopsy Line 19 3 Voluson 730Pro Basic User Manual 105831 Rev 0 19 1 ...
Страница 414: ...Biopsy Setup This page intentionally left blank Voluson 730Pro Basic User Manual 19 4 105831 Rev 0 ...
Страница 442: ...Probes and Biopsy This page intentionally left blank Voluson 730Pro Basic User Manual 20 28 105831 Rev 0 ...
Страница 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...
Страница 458: ...Connections 21 2 11 ECG preamplifier MAN Connection Voluson 730Pro Basic User Manual 21 16 105831 Rev 0 ...