System Setup
17.3.8 DICOM
DICOM is the abbreviation of Digital Imaging and Communications in Medicine. This is the industrial
standard for communication of images and other information between medical devices on the network.
Using the DICOM option, you can send or print images after connecting your ultrasound equipment
and PACS.
This dialog section is used to set up details of all of your DICOM target nodes (image servers).
Once you have set up a DICOM node properly you can select it without worrying about the details.
AE
(
A
pplication
E
ntity)
Title:
Please enter the Application Entity Title under which your
DICOM-application is known to other DICOM applications
(required). For setting the correct AE-Title please contact your
DICOM network administrator.
e.g.,
V730
Station Name:
Please enter the name of the hospital or institute.
Retry Count:
Number of retries to establish a non-successful DICOM connection.
Retry Interval:
Time pause in minutes between two attempts to establish a non-successful
DICOM
connection.
Test Connection:
Test the connection to a DICOM station.
First select the station that to test with the right or left trackball key, then click the
[
Test Connection
]
button. If the TCP/IP connection to the remote station is active, "OK" will appear in the
[
Ping
]
line. If
the DICOM server on the remote station is active, "OK" will appear in the
[
Verify
]
line.
This connection test may take up to 30 seconds.
Voluson
®
730Pro - Basic User Manual
105831 Rev. 0
17-19
Содержание Voluson 730Pro
Страница 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Страница 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Страница 22: ...CONTENTS This page intentionally left blank Voluson 730Pro Basic User Manual 18 CONTENTS 105831 Rev 0 ...
Страница 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Страница 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Страница 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Страница 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Страница 262: ...Volume Mode This page intentionally left blank Voluson 730Pro Basic User Manual 11 78 105831 Rev 0 ...
Страница 290: ...Basic Measurements This page intentionally left blank Voluson 730Pro Basic User Manual 13 16 105831 Rev 0 ...
Страница 411: ...Biopsy Setup 19 Biopsy Setup 19 2 19 1 To Program a Biopsy Line 19 3 Voluson 730Pro Basic User Manual 105831 Rev 0 19 1 ...
Страница 414: ...Biopsy Setup This page intentionally left blank Voluson 730Pro Basic User Manual 19 4 105831 Rev 0 ...
Страница 442: ...Probes and Biopsy This page intentionally left blank Voluson 730Pro Basic User Manual 20 28 105831 Rev 0 ...
Страница 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...
Страница 458: ...Connections 21 2 11 ECG preamplifier MAN Connection Voluson 730Pro Basic User Manual 21 16 105831 Rev 0 ...