Utilities
12.2 Histogram
With this function the gray scale or color distribution within a marked Region of Interest (ROI) will be
graphically displayed. Three histograms may be shown on the screen simultaneously.
There are three possibilities to calculate the gray scale or color distribution:
•
2D Histogram
(chapter 12.2.1)
•
3D Histogram
(chapter 12.2.2)
•
Volume Histogram
(chapter 12.2.3)
12.2.1 2D
Histogram
Operation:
1. Store a 2D mode, CFM mode or a PD mode image.
2. Switch on Histogram by pressing the
[
Utilities
]
key and selecting
[
Histogram
]
.
The menu area changes to the Histogram menu.
3. Select the number of histogram: 1, 2 or 3.
4. Use the trackball to place the rectangle over the ROI.
5. The upper trackball key changes from position to size of the
ROI and back.
6.
Press the right or left trackball key to calculate the histogram.
The histogram and corresponding number (left below the box) will be calculated and displayed.
Remarks:
*
Measuring, text annotations, bodymarks entering as well as all
post-processing settings are not possible in Histogram mode.
Display of gray scale HISTOGRAM
X-axis:
grayscale values from 0 to 255
Y-axis:
incidence in %, normalized to maximal incidence
A: Average
value
A = Sum of [values * presence]
Number
of
values
in
ROI
SD:
Standard
deviation
Voluson
®
730Pro - Basic User Manual
105831 Rev. 0
12-7
Содержание Voluson 730Pro
Страница 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Страница 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Страница 22: ...CONTENTS This page intentionally left blank Voluson 730Pro Basic User Manual 18 CONTENTS 105831 Rev 0 ...
Страница 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Страница 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Страница 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Страница 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Страница 262: ...Volume Mode This page intentionally left blank Voluson 730Pro Basic User Manual 11 78 105831 Rev 0 ...
Страница 290: ...Basic Measurements This page intentionally left blank Voluson 730Pro Basic User Manual 13 16 105831 Rev 0 ...
Страница 411: ...Biopsy Setup 19 Biopsy Setup 19 2 19 1 To Program a Biopsy Line 19 3 Voluson 730Pro Basic User Manual 105831 Rev 0 19 1 ...
Страница 414: ...Biopsy Setup This page intentionally left blank Voluson 730Pro Basic User Manual 19 4 105831 Rev 0 ...
Страница 442: ...Probes and Biopsy This page intentionally left blank Voluson 730Pro Basic User Manual 20 28 105831 Rev 0 ...
Страница 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...
Страница 458: ...Connections 21 2 11 ECG preamplifier MAN Connection Voluson 730Pro Basic User Manual 21 16 105831 Rev 0 ...