Probes and Biopsy
20.12.3 Disinfecting
Probes
Ultrasound probes can be disinfected using liquid chemical germicides. The level of disinfection is
directly related to the duration of contact with the germicide. Increased contact time produces a higher
level of disinfection.
Cidex and Sporox are the only germicides approved by GE Medical Systems for Ultrasound probe
disinfecting purposes. 2% Glutaraldehyde-based solutions (Cidex) have been shown to be very
effective for disinfecting.
Please consider our constantly updated Care-Card (which is inside the transducer boxes) for
compatible disinfectants and gels !
You also find this information on the web:
The Care website will be at
http://www.geultrasound.com/
and
http://www.gemedicalsystems.com/
To reach the Care and disinfectant listing site please select
Radiology -
Ultrasound - Probes!
In order for liquid chemical germicides to be effective, all visible residue must be
removed during the cleaning process. Thoroughly clean the probe, as described
earlier, before attempting disinfection.
1. Prepare the germicide solution according to the manufacturer's instructions. Be sure to follow all
precautions for storage, use and disposal.
2. Place the cleaned and dried probe in contact with the germicide for the time specified by the
germicide manufacturer. High-level disinfection is recommended for surface probes and is
required for endocavitary and intraoperative probes (follow the germicide manufacturer's
recommended time).
3. After removing from the germicide, rinse the probe following the germicide manufacturer's
rinsing instructions. Flush all visible germicide residue from the probe and allow to air dry.
CREUTZFIELD-JACOB DISEASE
Neurological use on patients with this disease must be avoided.
If a probe becomes contaminated, there is no adequate disinfecting means.
Voluson
®
730Pro - Basic User Manual
105831 Rev. 0
20-17
Содержание Voluson 730Pro
Страница 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Страница 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Страница 22: ...CONTENTS This page intentionally left blank Voluson 730Pro Basic User Manual 18 CONTENTS 105831 Rev 0 ...
Страница 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Страница 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Страница 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Страница 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Страница 262: ...Volume Mode This page intentionally left blank Voluson 730Pro Basic User Manual 11 78 105831 Rev 0 ...
Страница 290: ...Basic Measurements This page intentionally left blank Voluson 730Pro Basic User Manual 13 16 105831 Rev 0 ...
Страница 411: ...Biopsy Setup 19 Biopsy Setup 19 2 19 1 To Program a Biopsy Line 19 3 Voluson 730Pro Basic User Manual 105831 Rev 0 19 1 ...
Страница 414: ...Biopsy Setup This page intentionally left blank Voluson 730Pro Basic User Manual 19 4 105831 Rev 0 ...
Страница 442: ...Probes and Biopsy This page intentionally left blank Voluson 730Pro Basic User Manual 20 28 105831 Rev 0 ...
Страница 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...
Страница 458: ...Connections 21 2 11 ECG preamplifier MAN Connection Voluson 730Pro Basic User Manual 21 16 105831 Rev 0 ...