Spectral Doppler Mode
7.1.2.7 Invert
This function inverts the PW spectrum display in relation to the direction of the flow. The displayed
spectrum is inverted around the baseline. The velocity or frequency scale changes accordingly.
Use
[
Invert
]
when necessary to change the spectral display orientation.
It is possible in both read- and write mode.
Forward flow indicates: Blood flow toward the transducer
Reverse flow indicates: Blood flow away from the transducer
key unlit Normal:
Forward flow above the baseline
Reverse flow below the baseline
key lit:
Inverted:
Reverse flow above the baseline
Forward flow below the baseline
7.1.2.8 Angle
Correction
To get optimum resolution and accuracy from Doppler measurements, the angle that exists between
the ultrasound beam and the blood flow should be maintained between 0 and 20 degrees.
However, due to anatomical limitations an angle of 55 to 65 degrees is common in peripheral vascular
applications. The blood flow velocity calculation based on the incident angle of the ultrasound beam to
the axis of the vessel can be determined this way.
The vessel must be displayed in longitudinal section and the angle cursor must be positioned parallel
to the vessel axis (in the area of the measuring volume).
Angle correction adjusts the Doppler scale and is only necessary for velocity display (cm/s, m/s)
according to the Doppler equation.
The cursor angle can be changed in 1° increments in both directions continuously.
By
pressing
the angle knob repeatedly the angle correction is changed
from + 60° to 0° and to – 60°.
No indication to set the angle correction will appear in the measuring programs.
Remarks:
*
The actual angle is displayed on the screen [SV Angle ...].
*
Adjusting the angle is always possible, in write mode and read mode.
Voluson
®
730Pro - Basic User Manual
105831 Rev. 0
7-7
Содержание Voluson 730Pro
Страница 1: ... B BA AS SI IC C U US SE ER R M MA AN NU UA AL L GE Medical Systems Kretztechnik GmbH Co OHG ...
Страница 4: ...This page intentionally left blank Voluson 730Pro Operation Manual i 2 105831 Rev 0 ...
Страница 22: ...CONTENTS This page intentionally left blank Voluson 730Pro Basic User Manual 18 CONTENTS 105831 Rev 0 ...
Страница 23: ...General 1 General 1 2 Voluson 730Pro Basic User Manual 105831 Rev 0 1 1 ...
Страница 44: ...Safety This page intentionally left blank Voluson 730Pro Basic User Manual 2 20 105831 Rev 0 ...
Страница 112: ...2D Mode This page intentionally left blank Voluson 730Pro Basic User Manual 5 22 105831 Rev 0 ...
Страница 172: ...PD Mode This page intentionally left blank Voluson 730Pro Basic User Manual 9 14 105831 Rev 0 ...
Страница 262: ...Volume Mode This page intentionally left blank Voluson 730Pro Basic User Manual 11 78 105831 Rev 0 ...
Страница 290: ...Basic Measurements This page intentionally left blank Voluson 730Pro Basic User Manual 13 16 105831 Rev 0 ...
Страница 411: ...Biopsy Setup 19 Biopsy Setup 19 2 19 1 To Program a Biopsy Line 19 3 Voluson 730Pro Basic User Manual 105831 Rev 0 19 1 ...
Страница 414: ...Biopsy Setup This page intentionally left blank Voluson 730Pro Basic User Manual 19 4 105831 Rev 0 ...
Страница 442: ...Probes and Biopsy This page intentionally left blank Voluson 730Pro Basic User Manual 20 28 105831 Rev 0 ...
Страница 446: ...Connections 21 2 1 Main Module Voluson 730Pro Basic User Manual 21 4 105831 Rev 0 ...
Страница 458: ...Connections 21 2 11 ECG preamplifier MAN Connection Voluson 730Pro Basic User Manual 21 16 105831 Rev 0 ...