Site Requirements
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
3–5
Environmental Requirements
The system should be operated, stored, or transported within
the parameters outlined below. Either its operational
environment must be constantly maintained or the unit must be
turned off.
Operational
Storage
Transport (<16 hrs.)
Temperature
10
- 40
C
50
- 104
F
–10
- 60
C
14
- 140
F
–40
- 60
C
–40
- 140
F
Humidity
30-85%
non-condensing
30-90%
non-condensing
30-90%
non-condensing
Pressure
700-1060hPa
700-1060hPa
700-1060hPa
Table 3–1. System Environmental Requirements
Содержание 500
Страница 1: ... ...
Страница 58: ...Safety Precautions LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 2 28 This page left blank intentionally ...
Страница 104: ...Operator Controls LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 3 46 This page left blank intentionally ...
Страница 112: ...Exam Applic Preset Selection LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 4 8 This page left blank intentionally ...
Страница 262: ...General Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 8 12 This page left blank intentionally ...
Страница 312: ...GYN Summary Report LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 9 50 This page left blank intentionally ...
Страница 350: ...Advanced Cardiac Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 10 38 This page left blank intentionally ...
Страница 364: ...Urology Basic Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 12 4 This page left blank intentionally ...
Страница 404: ...Advanced Recording Option LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 13 40 This page left blank intentionally ...
Страница 610: ...Assistance LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 16 44 This page left blank intentionally ...