OB Measurements and Formulas
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
9–9
Osaka University Method (cont’d)
Calc
Mnemonic
Calc Name
Input Measurements
Formula
Author
Reference
EFBW
Estimated Fetal
Body Weight
Average of BPD, Average
of FTA and Average of FL
EFBW=1.25647x
BPD^3+3.50655
x FTA x FL+6.3 (<5000g)
IUGR=1.229 x
BPD^3+3.063 x
FTA x FL–24.6
Curve data is available in
the Advanced Reference
Manual
Osaka University
Method 1989,
3 by Univ. of
Osaka
CTAR
Cardio–Thoracic
Area Ratio
four distances or two
areas
CTAR = (d1 x d2) /
(d3 x d4) x 100
or a1 / a2 x 100
Y.Chiba Fetal
Diagn Ther 1990 :
5 : 175 – 188
PLI
Preload Index
two Doppler blood flow
velocities
PLI = | v1 / v2 |
Y.Chiba Fetal
Diagn Ther 1990 :
5 : 168 – 174
EF
Ejection Fraction
two distances on M–Mode
(End–diastolic dimension
and End–systolic
dimension on M–Mode)
EF = (1 – Ds^3 / Dd^3)
n/a
Vmax
Maximum Velocity
in descending
aorta
one Doppler blood flow
peak velocities
Vmax = input
n/a
S/D Ratio
Systolic / Diastolic
Ratio
two Doppler blood flow
peak velocities
S/D = Vsystole /
Vdiastole
*NOTE*
n/a
A/B Ratio
Velocities Ratio
two Doppler blood flow
peak velocities
A/B = V
1
/V
2
n/a
Pl
Pulsatility Index
two Doppler blood flow
peak velocities and
TA_MAX
Pl = (Vmax – V
diastole
) /
TA_MAX
*NOTE*
n/a
RI
Resistivity Index
two Doppler blood flow
peak velocities
RI = (Vmax – V
diastole
) /
Vmax
*NOTE*
n/a
*NOTE*:
V
diastole
= V
min
or V
end–diastole
(depends on preset selection)
Table 9–2. OB Calculation Formulas—Osaka University Factory Defaults (cont’d)
Содержание 500
Страница 1: ... ...
Страница 58: ...Safety Precautions LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 2 28 This page left blank intentionally ...
Страница 104: ...Operator Controls LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 3 46 This page left blank intentionally ...
Страница 112: ...Exam Applic Preset Selection LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 4 8 This page left blank intentionally ...
Страница 262: ...General Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 8 12 This page left blank intentionally ...
Страница 312: ...GYN Summary Report LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 9 50 This page left blank intentionally ...
Страница 350: ...Advanced Cardiac Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 10 38 This page left blank intentionally ...
Страница 364: ...Urology Basic Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 12 4 This page left blank intentionally ...
Страница 404: ...Advanced Recording Option LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 13 40 This page left blank intentionally ...
Страница 610: ...Assistance LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 16 44 This page left blank intentionally ...