LV Measurement Methods
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
10–10
Teichholz Method (cont’d)
Calc Mnemonic
Calc Name
Input Measurements
Formula
LVIDd
Left Ventricular Internal
Diameter, Diastole
one distance
LVIDd=d1[cm or mm]
LVIDs
Left Ventricular Internal
Diameter, Systole
one distance
LVIDs=d1[cm or mm]
IVSd
Interventricular Septal
Thickness, Diastole
one distance
IVSd=d1[cm or mm]
IVSs
Interventricular Septal
Thickness, Systole
one distance
IVSs=d1[cm or mm]
LVPWd
Left Ventricle Posterior
Wall Thickness, Diastole
one distance
LVPWd=d1[cm or mm]
LVPWs
Left Ventricle Posterior
Wall Thickness, Systole
one distance
LVPWs=d1[cm or mm]
EdV
End Diastole Volume
one distance
EdV[ml]=LVIDd^3x7/(2.4
+LVIDd)
EsV
End Systole Volume
one distance
EsV[ml]=LVIDs^3x7/(2.4
+LVIDs)
SV
Stroke Volume
two distances
SV[ml]=EdV–EsV
CO
Cardiac Output
two distances and one 2
beat time interval
CO[1/min]=
SVxHR/1000
EF
Ejection Fraction
two distances
EF=SV/EdVx100
FS
Fractional Shortening
two distances
FS=(1–LVIDs/
LVIDd)x100
HR
Heart Rate
(beats/minute)
one 2 beat time
interval
HR=120[sec]/
2 beat time[sec]
PHT
Pressure Half Time
one time interval
PHT=t1[ms orsec]
MVA
Mitral Valve Area
one pressure half time
interval
MVA[cm^2]= 220/PHT
ET
Ejection Time
one time interval
ET=t1[ms or sec]
Table 10–2. Teichholz Method Formulas
Содержание 500
Страница 1: ... ...
Страница 58: ...Safety Precautions LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 2 28 This page left blank intentionally ...
Страница 104: ...Operator Controls LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 3 46 This page left blank intentionally ...
Страница 112: ...Exam Applic Preset Selection LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 4 8 This page left blank intentionally ...
Страница 262: ...General Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 8 12 This page left blank intentionally ...
Страница 312: ...GYN Summary Report LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 9 50 This page left blank intentionally ...
Страница 350: ...Advanced Cardiac Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 10 38 This page left blank intentionally ...
Страница 364: ...Urology Basic Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 12 4 This page left blank intentionally ...
Страница 404: ...Advanced Recording Option LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 13 40 This page left blank intentionally ...