B–Mode
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
5–23
B-Mode Sub-Menu Page 4
SGL
Preset
B
Edge
Enhance
Rejectn
Image
Rotatn
Biopsy
Zone
Set Up
ECG
0DEG
40
MID
Figure 5–9. B-Mode Sub-Menu Page 4
Biopsy Zone
Image Rotate
Description
This selection enables any electronic biopsy
guidezone(s) available for the active probe.
If the single fixed angle (SGL) was selected as
the needle guide type in Set Up/Custom
Display page 1, the fixed zone angle is
displayed.
If the needle guide type selected is multiple
angle (MBX), the Biopsy Zone rocker switch
can be used to cycle through the angle
selections MBX1, MBX2, MBX3 and OFF.
Rotates the single real-time or
zoomed B-Mode image in 90
increments.
When B-Mode is rotated 180
, the
M-Mode (if active) is also rotated
180
.
Benefits
The electronic biopsy guidezone shows the
expected needle path during insertion.
Failure to match the
guidezone displayed to
the guide may cause the
needle to track a path
outside the zone.
DANGER
Orient image display for easy
reference.
outside the zone.
It is extremely important that when using the
adjustable angle biopsy guides, the angle
displayed on the screen matches the angle set
on the guide, otherwise the needle will not
follow the displayed guidezone which could
result in repeated biopsies or patient injury.
Safety
Concern
When the biopsy
guidezone is displayed,
the message:
“Confirm BX type of Bracket”
is displayed at the bottom of the screen with the
angle selected
DANGER
angle selected.
Ensure that the (M)BX type selected for each
probe is the same as the angle selected on the
biopsy guide.
NOTE: For specifics on availability of biopsy
guide angles, see 15–39.
Содержание 500
Страница 1: ... ...
Страница 58: ...Safety Precautions LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 2 28 This page left blank intentionally ...
Страница 104: ...Operator Controls LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 3 46 This page left blank intentionally ...
Страница 112: ...Exam Applic Preset Selection LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 4 8 This page left blank intentionally ...
Страница 262: ...General Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 8 12 This page left blank intentionally ...
Страница 312: ...GYN Summary Report LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 9 50 This page left blank intentionally ...
Страница 350: ...Advanced Cardiac Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 10 38 This page left blank intentionally ...
Страница 364: ...Urology Basic Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 12 4 This page left blank intentionally ...
Страница 404: ...Advanced Recording Option LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 13 40 This page left blank intentionally ...
Страница 610: ...Assistance LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 16 44 This page left blank intentionally ...