3D–Surface Mode
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
5–82
3D-Surface Mode
(Option)
Overview
This option creates a projected image different from 3DvieW. It
uses images stored in Cine. The calculation used for the
projection is based on the transmission principle of light using
Volume-Rendering.
This function is available only in single B-Mode.
Refer to 3DvieW Mode section as needed.
This 3D-Surface option is designed for qualitative information
only and has no provision for any measurement or computation
ability.
No specific reconstruction resolution or accuracy is provided or
claimed for this option.
Since positional information of the original scan is not available,
all measurement functions cannot be performed on a
3D-Surface image.
NOTE: Consider all of the above cautions when using
3D-Surface images to evaluate a patient. 3D-Surface images
should not be used for quantitative information but a qualitative
supplement to all other diagnostic data collected for that patient.
Refer to the
Advanced Reference Manual
for further details.
CAUTION
CAUTION
Содержание 500
Страница 1: ... ...
Страница 58: ...Safety Precautions LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 2 28 This page left blank intentionally ...
Страница 104: ...Operator Controls LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 3 46 This page left blank intentionally ...
Страница 112: ...Exam Applic Preset Selection LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 4 8 This page left blank intentionally ...
Страница 262: ...General Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 8 12 This page left blank intentionally ...
Страница 312: ...GYN Summary Report LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 9 50 This page left blank intentionally ...
Страница 350: ...Advanced Cardiac Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 10 38 This page left blank intentionally ...
Страница 364: ...Urology Basic Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 12 4 This page left blank intentionally ...
Страница 404: ...Advanced Recording Option LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 13 40 This page left blank intentionally ...
Страница 610: ...Assistance LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 16 44 This page left blank intentionally ...