Additional Cardiac Calculations
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
10–25
ET (Ejection Time)
Ejection time for the left ventricle is the time during which the
aortic valve is open. It is measured on a Doppler Spectrum
display. The time between the two cursors is ejection time (ET).
Figure 10–28. Ejection Time for Aortic Valve
Select ET from the Calculation Sub-Menu. Perform a time
interval measurement as described on
7–18
. Set the first
cursor at the opening of the aortic valve and the second at the
closing of the aortic valve.
MVA
Mean PG
LV
Report
PHT
ET
Max PG
TRACE
AUTO
Figure 10–29. Cardiac Calculation Sub-Menu
Содержание 500
Страница 1: ... ...
Страница 58: ...Safety Precautions LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 2 28 This page left blank intentionally ...
Страница 104: ...Operator Controls LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 3 46 This page left blank intentionally ...
Страница 112: ...Exam Applic Preset Selection LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 4 8 This page left blank intentionally ...
Страница 262: ...General Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 8 12 This page left blank intentionally ...
Страница 312: ...GYN Summary Report LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 9 50 This page left blank intentionally ...
Страница 350: ...Advanced Cardiac Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 10 38 This page left blank intentionally ...
Страница 364: ...Urology Basic Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 12 4 This page left blank intentionally ...
Страница 404: ...Advanced Recording Option LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 13 40 This page left blank intentionally ...
Страница 610: ...Assistance LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 16 44 This page left blank intentionally ...