OB Measurements and Formulas
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
9–7
Tokyo University Method (cont’d)
Calc
Mnemonic
Calc Name
Input Measurements
Formula
Author
Reference
HR
Heart Rate
(beats/minute)
one 2 beat time
(measured manually or
automatically)
HR[BPM]=
120[sec]/
2beat time[sec]
n/a
GS
Gestational Sac
one distance
Curve data is available
in the Advanced
Tokyo University
Method 1986
CRL
Crown Rump
Length
one distance
in the Advanced
Reference Manual
Method 1986,
6 by Univ. of
Tokyo
FL
Femur Length
one distance
y
BPD
Biparietal
Diameter
one distance
LV
Length of Vertebra
one distance
APTD
Antero posterior
Trunk Diameter
one distance
APTD = input
TTD
Transverse Trunk
Diameter
one distance
TTD = input
EFBW
Estimated Fetal
Body Weight
Average of BPD, Average
of APTD, Average of TTD
and Average of FL
EFBW = 1.07 x
BPD^ 3 +3.42 x APTD
x TTD x FL
CTAR
Cardio–Thoracic
Area Ratio
four distances or two
areas
CTAR =
(d1 x d2) / (d3 x d4) x
100 or a1 / a2 x 100
Y.Chiba Fetal
Diagn Ther 1990 :
5 : 175 – 188
PLI
Preload Index
two Doppler blood flow
velocities
PLI = | v1 / v2 |
Y.Chiba Fetal
Diagn Ther 1990 :
5 : 168 – 174
EF
Ejection Fraction
two distances on M-Mode
(End-diastolic dimension
and End-systolic
dimension on M-Mode)
EF = (1 – Ds^3 / Dd^3)
n/a
Vmax
Maximum Velocity
in descending
aorta
one Doppler blood flow
peak velocities
Vmax = input
n/a
S/D Ratio
Systolic / Diastolic
Ratio
two Doppler blood flow
peak velocities
S/D = Vsystole /
Vdiastole
*NOTE*
n/a
A/B Ratio
Velocities Ratio
two Doppler blood flow
peak velocities
A/B = V
1
/V
2
n/a
Pl
Pulsatility Index
two Doppler blood flow
peak velocities and
TA_MAX
Pl = (Vmax – V
diastole
)/
TA_MAX
*NOTE*
n/a
RI
Resistivity Index
two Doppler blood flow
peak velocities
RI = (Vmax – V
diastole
)/
Vmax
*NOTE*
n/a
*NOTE*:
V
diastole
= V
min
or V
end–diastole
(depends on preset selection)
Table 9–1. OB Calculation Formulas—Tokyo University Factory Defaults
Содержание 500
Страница 1: ... ...
Страница 58: ...Safety Precautions LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 2 28 This page left blank intentionally ...
Страница 104: ...Operator Controls LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 3 46 This page left blank intentionally ...
Страница 112: ...Exam Applic Preset Selection LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 4 8 This page left blank intentionally ...
Страница 262: ...General Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 8 12 This page left blank intentionally ...
Страница 312: ...GYN Summary Report LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 9 50 This page left blank intentionally ...
Страница 350: ...Advanced Cardiac Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 10 38 This page left blank intentionally ...
Страница 364: ...Urology Basic Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 12 4 This page left blank intentionally ...
Страница 404: ...Advanced Recording Option LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 13 40 This page left blank intentionally ...
Страница 610: ...Assistance LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 16 44 This page left blank intentionally ...