Mode Measurements
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
7–22
Depth Difference with Time and Slope
To measure depth difference:
Press Measurement 3 times. With the cursor in the M-Mode
timeline area, a “
” cursor with a vertical dotted line is
displayed.
Use the Trackball to position the measurement start point.
Press Set to fix the start point.
Use the Trackball to position the measurement end point.
Press Set to complete the measurement.
The depth difference, time interval and slope between the 2 end
points is displayed at the bottom of the screen.
Содержание 500
Страница 1: ... ...
Страница 58: ...Safety Precautions LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 2 28 This page left blank intentionally ...
Страница 104: ...Operator Controls LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 3 46 This page left blank intentionally ...
Страница 112: ...Exam Applic Preset Selection LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 4 8 This page left blank intentionally ...
Страница 262: ...General Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 8 12 This page left blank intentionally ...
Страница 312: ...GYN Summary Report LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 9 50 This page left blank intentionally ...
Страница 350: ...Advanced Cardiac Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 10 38 This page left blank intentionally ...
Страница 364: ...Urology Basic Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 12 4 This page left blank intentionally ...
Страница 404: ...Advanced Recording Option LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 13 40 This page left blank intentionally ...
Страница 610: ...Assistance LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 16 44 This page left blank intentionally ...