Introduction
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
7–2
Introduction
Overview
Measurements and calculations derived from ultrasound
images are intended to supplement other clinical procedures
available to the attending physician. The accuracy of
measurements is not only determined by system accuracy, but
also by the use of proper medical protocols by the user. When
appropriate, be sure to note any protocols associated with a
particular measurement or calculation. Formulas and
databases used within the system software that are associated
with specific investigators are so noted. Be sure to refer to the
original article describing the investigator’s recommended
clinical procedures.
Measurements can be made in all modes and image formats,
including real-time, freeze, CINE or VCR playback. The
selected exam category defines the default calculations
displayed on the Softkey Sub-Menu.
Measurement values update on the display until the
measurement has been completed. 18 lines of measurements
can be displayed on the screen at one time.
The measurement cursor is not automatically fixed when the
Record key is pressed. Record all desired measurements
displayed on the screen before erasing them.
Содержание 500
Страница 1: ... ...
Страница 58: ...Safety Precautions LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 2 28 This page left blank intentionally ...
Страница 104: ...Operator Controls LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 3 46 This page left blank intentionally ...
Страница 112: ...Exam Applic Preset Selection LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 4 8 This page left blank intentionally ...
Страница 262: ...General Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 8 12 This page left blank intentionally ...
Страница 312: ...GYN Summary Report LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 9 50 This page left blank intentionally ...
Страница 350: ...Advanced Cardiac Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 10 38 This page left blank intentionally ...
Страница 364: ...Urology Basic Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 12 4 This page left blank intentionally ...
Страница 404: ...Advanced Recording Option LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 13 40 This page left blank intentionally ...
Страница 610: ...Assistance LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 16 44 This page left blank intentionally ...