GYN Measurements
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
9–46
Resistive index
RI measurements are velocity values measured on the Doppler
Spectrum for vessels in the ovaries and uterus.
To make GYN Resistive Index measurements:
Scan the desired vessel in Doppler mode.
Select the desired RI measurement from the GYN Calculation
Sub-Menu.
LtOvFo
RtOvFo
Ov Fol
Report
Figure 9–17. GYN Calculation Sub-Menu
Use the Trackball to move the cursor to the peak point of the
Doppler waveform where it is desired to start the RI
measurement.
Press Set to fix the start-point cursor. An end-point cursor
appears.
Use the Trackball to move the end-point cursor to the diastolic
point of the Doppler waveform.
Press Set to complete the measurement. A calculated value is
displayed and stored in the GYN Summary Report.
Содержание 500
Страница 1: ... ...
Страница 58: ...Safety Precautions LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 2 28 This page left blank intentionally ...
Страница 104: ...Operator Controls LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 3 46 This page left blank intentionally ...
Страница 112: ...Exam Applic Preset Selection LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 4 8 This page left blank intentionally ...
Страница 262: ...General Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 8 12 This page left blank intentionally ...
Страница 312: ...GYN Summary Report LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 9 50 This page left blank intentionally ...
Страница 350: ...Advanced Cardiac Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 10 38 This page left blank intentionally ...
Страница 364: ...Urology Basic Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 12 4 This page left blank intentionally ...
Страница 404: ...Advanced Recording Option LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 13 40 This page left blank intentionally ...
Страница 610: ...Assistance LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 16 44 This page left blank intentionally ...