Custom Display
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
14–16
Page 2 of 20 (Imaging Parameter 2 – Probe Dependent 2)
B Gain EC, UP, P, R
Choose the default gain value for B-Mode.
Value from 0 to 98 in 2 digit increments. The maximum value is probe dependent and
may not reach 98.
B Dynamic Range EC, UP, P, R
Choose the default dynamic range value for B-Mode.
Value from 30 to 90 in 6 digit increments.
B Edge Enhance EC, UP, P, R
Choose the default edge enhancement value for B-Mode.
Off, Low, Mid or High.
B Image Softener EC, UP, P, R
Choose the B-Mode image softener function default.
On or Off.
B Softener Level EC, UP, P, R
Choose the B-Mode image softener default level.
D1, D2, D3, Low, Mid or High.
B Frame Average EC, UP, P
Choose the B-Mode frame averaging default value.
Off (no frame averaging), 1, 2, 3, 4, 5 or 6.
B Imaging Frequency EC, UP, P, R
Choose the B-Mode image frequency optimization setting.
Resolution
Process image for best resolution.
Normal
Compromise between resolution and penetration.
Penetration
Process image for penetration.
Further Penetration
C358 Probe Only—Process image for more penetration
Harmonic1
3S, S222, S317 and C358 Probes Only—Process image with
tissue harmonic frequency.
Harmonic2
3S, S222, S317 and C358 Probes Only—Process image with
tissue harmonic frequency.
The soft menu display shows this parameter as an imaging frequency depending on the
active probe.
Содержание 500
Страница 1: ... ...
Страница 58: ...Safety Precautions LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 2 28 This page left blank intentionally ...
Страница 104: ...Operator Controls LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 3 46 This page left blank intentionally ...
Страница 112: ...Exam Applic Preset Selection LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 4 8 This page left blank intentionally ...
Страница 262: ...General Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 8 12 This page left blank intentionally ...
Страница 312: ...GYN Summary Report LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 9 50 This page left blank intentionally ...
Страница 350: ...Advanced Cardiac Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 10 38 This page left blank intentionally ...
Страница 364: ...Urology Basic Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 12 4 This page left blank intentionally ...
Страница 404: ...Advanced Recording Option LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 13 40 This page left blank intentionally ...
Страница 610: ...Assistance LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 16 44 This page left blank intentionally ...