Probe Overview
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
15–17
Disinfecting probes
Ultrasound probes can be disinfected using liquid chemical
germicides. The level of disinfection is directly related to the
duration of contact with the germicide. Increased contact time
produces a higher level of disinfection.
Cidex and Sporox are the only germicides approved by GE
Medical Systems for Ultrasound probe disinfecting purposes.
2% Glutaraldehyde-based solutions have been shown to be
very effective for disinfecting. Cidex (a 2% Glutaraldehyde-
based germicide) and Sporox (an alternative germicide with
non-glutaraldehyde-based chemicals) have been evaluated for
compatibility with the material used to construct the probes.
The following FDA cleared, high-level disinfectant chemicals
are recommended:
Cidex—All Probes.
Cidex OPA—All Probes,
EXCEPT 348C and 548C on
LOGIQ
700 systems
.
Sporox—All Probes,
EXCEPT 348C & 548C on
LOGIQ
700 systems AND TE (Trans esophageal)
.
Cidex PA—TE (Trans esophageal) Probes ONLY.
In order for liquid chemical germicides to be effective, all visible
residue must be removed during the cleaning process.
Thoroughly clean the probe, as described earlier before
attempting disinfection.
1.
Prepare the germicide solution according to the
manufacturer’s instructions. Be sure to follow all
precautions for storage, use and disposal.
After Each Use
WARNING
CAUTION
Содержание 500
Страница 1: ... ...
Страница 58: ...Safety Precautions LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 2 28 This page left blank intentionally ...
Страница 104: ...Operator Controls LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 3 46 This page left blank intentionally ...
Страница 112: ...Exam Applic Preset Selection LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 4 8 This page left blank intentionally ...
Страница 262: ...General Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 8 12 This page left blank intentionally ...
Страница 312: ...GYN Summary Report LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 9 50 This page left blank intentionally ...
Страница 350: ...Advanced Cardiac Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 10 38 This page left blank intentionally ...
Страница 364: ...Urology Basic Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 12 4 This page left blank intentionally ...
Страница 404: ...Advanced Recording Option LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 13 40 This page left blank intentionally ...
Страница 610: ...Assistance LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 16 44 This page left blank intentionally ...