Custom Display
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
14–25
Page 6 of 20 (Imaging Parameter 6 – Probe Dependent 6) (cont’d)
CFM Velocity [cm/s] EC, UP, P, R
Choose the default CFM velocity scale in centimeters per second.
Value from 0 to 1000 in 10 digit increments.
CFM B Gain EC, UP, P, R
Choose the default B-Mode gain setting while in CFM Mode.
Value from 0 to 31 in 1 digit increments.
CFM B High Resolution EC, UP, P, R
Choose the initial default setting of the high resolution function in CFM Mode.
On or Off.
CFM B Frame Average EC, UP, P, R
Choose the level of frame averaging for CFM while in combination B-Mode with CFM.
Off
No B-Mode frame averaging.
Low
Low frame averaging.
Mid
Medium frame averaging.
High
High frame averaging.
CFM B Wall Filter EC, UP, P, R
Choose the default level of wall filtering for B-Mode while in CFM.
Low, M1, M2, M3, M4 and High.
CFM B Angle/Width [deg, mm] EC, UP, P, R
Choose the default B-Mode angle or width of the display while in CFM.
Value from 1 to 124 in one digit increments. Maximum is determined by the probe type.
CFM/PDI Start Depth [mm] EC, UP, P, R
Set the start depth of the CFM/PDI window area.
Depth in millimeters ranging from 0 to 230.
CFM/PDI Vertical Size [mm] EC, UP, P, R
Set the vertical size of the CFM/PDI window.
Size in millimeters ranging from 10 to 240.
CFM/PDI Focus Depth [%] EC, UP, P, R
Set the desired depth of the focal zone in CFM/PDI Mode window.
Percentage ranging from 0 to 100.
Содержание 500
Страница 1: ... ...
Страница 58: ...Safety Precautions LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 2 28 This page left blank intentionally ...
Страница 104: ...Operator Controls LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 3 46 This page left blank intentionally ...
Страница 112: ...Exam Applic Preset Selection LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 4 8 This page left blank intentionally ...
Страница 262: ...General Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 8 12 This page left blank intentionally ...
Страница 312: ...GYN Summary Report LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 9 50 This page left blank intentionally ...
Страница 350: ...Advanced Cardiac Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 10 38 This page left blank intentionally ...
Страница 364: ...Urology Basic Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 12 4 This page left blank intentionally ...
Страница 404: ...Advanced Recording Option LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 13 40 This page left blank intentionally ...
Страница 610: ...Assistance LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 16 44 This page left blank intentionally ...