GYN Summary Report
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
9–49
GYN Calculation Formulas
Calc
Mnemonic
Calc Name
Input
Measurements
Formula
UT-L
Uterine Length
one distance
Ut-L[cm or mm]=d1
UT-H
Uterine Height
one distance
Ut-H[cm or mm]=d1
UT-W
Uterine Width
one distance
Ut-W[cm or mm]=d1
Endo
Endometrium Thick-
ness
one distance
Endo[cm or mm]=d1
Lt. Ov-L
Left Ovarian Length
one distance
Lt. Ov-L[cm or mm]=d1
Lt. Ov-H
Left Ovarian Height
one distance
Lt. Ov-H[cm or mm]=d1
Lt. Ov-W
Left Ovarian Width
one distance
Lt. Ov-W[cm or mm]=d1
Rt. Ov-L
Right Ovarian Length
one distance
Rt. Ov-L[cm or mm]=d1
Rt. Ov-H
Right Ovarian Height
one distance
Rt. Ov-H[cm or mm]=d1
Rt. Ov-W
Right Ovarian Width
one distance
Rt. Ov-W[cm or mm]=d1
Lt. Ov-RI
Left Ovarian Vessel
Resistive Index
two Doppler blood flow
peak velocities
Lt. Ov-RI= (V
max
–V
diastole
)/V
max
*NOTE*
Ut-RI
Uterine Vessel Resis-
tive Index
two Doppler blood flow
peak velocities
Ut-RI= (V
max
–V
diastole
)/V
max
*NOTE*
Rt. Ov-RI
Right Ovarian Vessel
Resistive Index
two Doppler blood flow
peak velocities
Rt. Ov-RI= (V
max
–V
diastole
)/V
max
*NOTE*
LtOvFo[ml]
Left Ovary Follicles
One distance*
D1[cm]
3
x
p
/6
Two distances*
D1[cm]
2
x D2x[cm]
p
/6
: (D1 <D2)
D1[cm] x D2[cm]
2
x
p
/6
: (D2 <D1)
Three distances
D1[cm] x D2[cm]x D3[cm] x
p
/6
RtOvFo[ml]
Rith Ovary Follicles
One distance*
D1[cm]
3
x
p
/6
Two distances*
D1[cm]
2
x D2[cm] x
p
/6
: (D1 <D2)
D1[cm] x D2[cm]
2
x
p
/6
: (D2 <D1)
Three distances
D1[cm] x D2[cm] x D3[cm] x
p
/6
*NOTE*:
V
diastole
= V
min
or V
end–diastole
(depends on preset selection)
* To calculate LtOvFo or RtOvFo with one (or two distances), press the Clear key
after the first (or second distance) measurement(s).
Table 9–8. GYN Calculation Formulas
Содержание 500
Страница 1: ... ...
Страница 58: ...Safety Precautions LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 2 28 This page left blank intentionally ...
Страница 104: ...Operator Controls LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 3 46 This page left blank intentionally ...
Страница 112: ...Exam Applic Preset Selection LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 4 8 This page left blank intentionally ...
Страница 262: ...General Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 8 12 This page left blank intentionally ...
Страница 312: ...GYN Summary Report LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 9 50 This page left blank intentionally ...
Страница 350: ...Advanced Cardiac Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 10 38 This page left blank intentionally ...
Страница 364: ...Urology Basic Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 12 4 This page left blank intentionally ...
Страница 404: ...Advanced Recording Option LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 13 40 This page left blank intentionally ...
Страница 610: ...Assistance LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 16 44 This page left blank intentionally ...