General Calculations
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
8–11
General Calculation Formulas
Calc
Mnemonic
Calc Name
Input Measurements
Formula
MaxPG
Maximum Pressure
Gradient
two Doppler blood flow peak
velocities
MaxPG[mmHg]=4x(v1^2–v2^2)
MeanPG
Mean Pressure Gradient
flow velocities from one time
marker to another time marker
in a Doppler display
MeanPG[mmHg]=
n
4x
ȍ
(Vi^2)/n
i=1
SV
Stroke Volume
flow velocities from one time
marker to another time marker
in a Doppler display and one
area
SV[ml]= (sum{V
t^2
} from t1 to
t2)x a1
CO
Cardiac Output
two distances and one 2 beat
time interval
CO[1/min]=
SVxHR/1000
FV
Flow Volume
flow velocities from one time
marker to another time marker
in a Doppler display and one
area
FV [ml] = (sum{Vt[m/s]} from
t1 to t2) x a1[cm^2]
FVO
Flow Volume Output
flow velocities from one time
marker to another time marker
in a Doppler display and one
area
FVO [ml/min] = FV [ml] /
(t2–t1) x 60000
% Stenosis
Stenosis Ratio
two areas (by ellipse, trace,
circle or distance)
% Stenosis= [1–(A
residual
/
A
lumen
)]x100
PI
Pulsatility Index
two Doppler blood flow peak
velocities and TAMAX
PI=(V
max
–V
diastole
)/TAMAX
*NOTE*
RI
Resistivity Index
two Doppler blood flow peak
velocities
RI=(V
max
–V
diastole
)/V
max
*NOTE*
HR
Heart Rate (beats/minute)
one 2 beat time interval
HR[BPM]=120[sec]/
2beat time [sec]
A/B Ratio
Velocities Ratio
two Doppler blood flow peak
velocities
A/B=V
1
/V
2
TAMAX
Time Averaged Maximum
Velocity (Trace Method is
Peak or manual)
two time marks in a Doppler
display
TAMAX=sum{V
t
} from t1 to
t2/(t2–t1) [cm/s or m/s]
TAMIN
Time Averaged Minimum
Velocity (Trace method is
Floor)
two time marks in a Doppler
display
TAMIN=sum{V
t
} from t1 to
t2/(t2–t1) [cm/s or m/s]
TAMEAN
Time Averaged Mean
Velocity (Trace method is
Mean)
two time marks in a Doppler
display
TAMEAN=sum{V
t
} from t1 to
t2/(t2–t1) [cm/s or m/s]
TAMODE
Time Averaged Mode
Velocity (Trace method is
Mode)
two time marks in a Doppler
display
TAMODE=sum{V
t
} from t1 to
t2/(t2–t1) [cm/s or m/s]
*NOTE*:
V
diastole
= V
min
or V
end–diastole
(depends on preset selection)
Table 8–1. General Calculation Formulas
Содержание 500
Страница 1: ... ...
Страница 58: ...Safety Precautions LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 2 28 This page left blank intentionally ...
Страница 104: ...Operator Controls LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 3 46 This page left blank intentionally ...
Страница 112: ...Exam Applic Preset Selection LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 4 8 This page left blank intentionally ...
Страница 262: ...General Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 8 12 This page left blank intentionally ...
Страница 312: ...GYN Summary Report LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 9 50 This page left blank intentionally ...
Страница 350: ...Advanced Cardiac Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 10 38 This page left blank intentionally ...
Страница 364: ...Urology Basic Calculations LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 12 4 This page left blank intentionally ...
Страница 404: ...Advanced Recording Option LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 13 40 This page left blank intentionally ...
Страница 610: ...Assistance LOGIQ 500 Basic Users Manual 2276612 100 Rev 0 16 44 This page left blank intentionally ...