M
ESSAGES
VF6
Infinity Vista
5-17
Message: Cannot
Analyze
ST<lead1> and
ST<lead2> parameter
value: <blanK>
ST cannot analyze with a paced
beat
• ST segment analysis cannot be
performed on pacemaker patients
Message: ST Too High
ST<lead1> and
ST<lead2> parameter
value: +++
Message: ST Too Low
ST<lead1> and
ST<lead2> parameter
value: ---
The detected ST value is outside
the measurement range.
• Check the patient and apply
treatment if necessary.
• Check the electrode placement
and change their position if
necessary.
Message: ST > UL
ST<lead1> and
ST<lead2> parameter
value: <value>
Message: ST < LL
ST<lead1> and
ST<lead2> parameter
value: <value>
The ST rate falls outside the
current upper or lower alarm limits.
The current alarm limits are
inappropriate for this patient.
• Observe the patient carefully.
• Apply treatment if clinically
indicated.
• Change the alarm limits as
described in the “Cardiac
Monitoring” chapter.
Message: ST Lead
<1> and/or ST Lead
<2> Invalid
ST<lead1> and/or
ST<lead2> parameter
value: <blank>
Invalid ST<leadname>.
ST<lead1> or ST<lead2>
connection restored from a fault.
Excessive signal noise.
Bad contact of electrodes.
Interference from auxiliary
equipment.
• Select correct ST lead(s).
• Verify ST lead(s) connection.
Heart Rate, Arrhythmia, ST Segment Analysis
(continued)
Message
Possible Cause
Suggested Action
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...