S
ELECTING
AND
P
REPARING
THE
T
RANSDUCER
VF6
Infinity Vista
14-7
3. Close the transducer stopcock to the patient.
4. Open the venting stopcock to air (atmosphere). The monitor displays a flat
waveform and a static IBP condition for the systolic (S) value.
5. Click on
Zero
.
6. Verify that the zero has been established. If zeroing failed, repeat steps 2 to 6.
If zeroing is successful, continue.
7. If applicable, select
Cal. Factor
and click the knob.
8. Dial in the calibration factor and click the knob.
9. If you are monitoring two invasive blood pressures, repeat these steps for the
second pressure parameter.
Zero and Calibration Check Troubleshooting
NOTE:
The monitor displays an error message (
IBP1 Cannot Zero
or
IBP2 Cannot
Zero
) when zeroing fails. Call your Biomed or replace the transducer if zeroing fails
after two attempts. If the monitor rejects the calibration factor that you selected, change
the transducer or call your Biomed. The transducer needs to be calibrated.
Screen Message
Possible Cause
Suggested Action
IBP Cannot Zero,
IBP Zero Time-out
• The transducer offset is outside
the zero balance range of +/-
190 mmHg
• The signal is too noisy.
• The waveform is non static
(more than 3 mmHg variation in
3 seconds).
• The monitor was unable to zero
the transducer within 10
seconds.
• Verify the pressure cable connection
to the monitor.
• Keep all tubing motionless.
• Check the stopcock and verify that the
system is open to air completely.
• Repeat the zeroing procedure.
• Replace the transducer if faulty.
IBP Invalid Cal
• The calibration factor is outside
the range of 80 to 120
(inclusive).
• Replace the transducer if calibration
fails after two attempts.
• Call your Biomed to calibrate the
transducer with a mercury manometer
if calibration fails repeatedly.
• Check the pressure cable for damage.
Note that instead of “IBP” the messages show the selected pressure label
(e.g., “
ART Cannot Zero
”).
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...