ECG M
ONITORING
S
ETTINGS
VF6
Infinity Vista
8-13
Pacer Detection
Pacer detection is used for paced patients in the adult and pediatric monitoring modes.
The monitor detects paced pulses in the amplitude range of ±5 to ±700 mV. When
pacer detection is enabled, the letter “P” is added to the flashing
j
symbol in the HR
parameter box for each paced beat. In addition, green positive spikes on the ECG
waveform indicate pacemaker pulses.
The monitor has an advanced pacer detection mode called
Fusion
, which offers
increased detection sensitivity to fused paced beats. To select the Pacer Fusion mode
in the HR menu, advanced pacer functions have to be enabled in the Biomed menu.
When monitoring paced patients, Dräger suggests the following:
z
Use the electrode positions that are best suited for paced patients (see the
electrode placement section).
z
Select the lead with the highest R-wave.
z
Follow the pacemaker precautions given at the end of this section (see
ECG
and HR Safety Considerations
).
STEPS: Selecting the Pacer Detection Mode
1. Select the
HR
parameter box and click the knob.
2. Click on
Pacer Detect
.
3. Dial in the desired setting (
ON, OFF or Fusion
) and click the knob.
NOTE:
The monitor indicates
Pacer Off
or
Pacer Fusion Mode
between the first two
waveform channels.
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...