D
EFAULT
S
ETTINGS
VF6
Infinity Vista
C-3
Adult
Settings
Pediatric
Settings
Neonatal
Settings
After patient discharge,
settings return to...?
ECG
Alarm
On
On
On
Saved Setup
Alarm Limits
(upper/lower)
120/45
150/50
180/80
Saved Setup
Alarm Recording
Off
Off
Off
Saved Setup
Tone Source
ECG
ECG
ECG
Previous Patient
Tone Volume
Off
Off
Off
Previous Patient
Pacer Detect
Off
Off
N/A
Saved Setup
QRS Marks
Off
Off
Off
Factory
ECG Processing
ECG 1&2
ECG 1&2
ECG 1
Previous Patient
ECG Leads
5-Lead
5-Lead
3-Lead
Previous Patient
Arrhythmia
*
ASY, VF Alarms
Always On
Always On
n/a
Previous Patient
VT, RUN, AIVR,
SVT, CPT, BGM
Alarm
BRDY, TACH,
PAUS, ARTF Alarm
On
Off
On
Off
n/a
n/a
Saved Setup
ASY and VF Alarm
Recording
Other Events
Record
Off
Record
Off
n/a
n/a
Saved Setup
Arrhythmia
Monitoring
Basic
Basic
n/a
Saved Setup
*Not applicable for neonatal monitoring.
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...