S
P
O2 M
ONITORING
S
ETTINGS
VF6
Infinity Vista
12-11
Signal Strength Bar Graph
The monitor can display a pulsing bar graph. The number of
illuminated segments is proportional to the pulse amplitude. Proper
sensor placement and environmental conditions ensure a strong
signal. Refer to the section
Sensor Application
for details.
STEPS: Displaying the Signal Strength Bar Graph
1. Click on the
SpO2
parameter box.
2. Click on
Bar Graph
.
3. Select
ON
or
OFF
and click the knob.
Averaging Mode
The monitor calculates the oxygen saturation of the arterial blood and the derived
pulse rate based on the averaging mode you select in the SpO
2
menu. The two modes
are Normal and Fast.
z
The Normal averaging mode updates the SpO
2
value and the derived pulse
rate in 30 seconds or less (8 seconds or less with the Masimo SET pod).
z
The Fast averaging mode updates the SpO
2
value and the derived pulse rate in
12 seconds or less (2 to 4 seconds with the Masimo SET pod).
Response time may improve as pulse rate increases.
NOTE:
The SpO
2
waveform and bar graph are auto-scaled by the Masimo SpO
2
pod.
NOTE:
When monitoring most patients, Dräger recommends using the Normal
averaging mode. The Fast mode is designed for neonatal patients when fast reporting of
oxygen desaturation is important.
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...