6-2
Infinity Vista
VF6
6 T
RENDS
Overview
The monitor can store the last 24 hours of trended data. If there is no alarm, the
monitor averages the monitored vital signs every minute and stores the results in its
trend memory.
You can view the trends in the following formats:
z
As a graph (graphical trends).
z
As a table (tabular trends).
Graphical and tabular trended values are displayed from left to right in increasing
time. If you discharge or admit a patient, the monitor erases trend data, patient data,
and stored recordings/events. Network time changes are reflected in both the graphical
and tabular trends.
By default, HR, SpO2 and NBP are the first three parameters displayed on the
graphical and tabular trend screens. However, you can select any three parameters for
the top positions on these screens (see below). The order of these parameters is saved
as part of the setup, when you save the current monitoring configuration (see the
chapter
Monitor Setup
).
For non-invasive blood pressure, the monitor stores a trend value at the end of a
successful measurement, or an error code at the end of an unsuccessful measurement,
but never more frequently than once every sixty seconds (i.e., if you take two NBP
measurements within one minute, the monitor stores only the last measurement as a
trend value).
Trend Setup
Select trend parameters for the three top positions on the trend display as follows:
STEPS: Selecting Parameters for Trend Channel Display
1. Press the
Menu
fixed key.
2. Click on
Monitor Setup
.
3. Click on
Trend Setup
.
4. Click on the desired trend display
Channel
.
5. Select the desired parameter and click the knob.
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...