O
VERVIEW
VF6
Infinity Vista
D-13
Interval Times
2, 2.5, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 30, 45, 60, 120,
180 and 240 min
Measuring Range
(Adult - 270mmHg)
Heart rate:
30 - 240 bpm
Systolic NBP:
30 - 250 mmHg
Mean NBP:
20 - 230 mmHg
Diastolic NBP:
10 - 210 mmHg
Measuring Range
(Pediatric - 180mmHg)
Heart rate:
30 - 240 bpm
Systolic NBP:
30 - 170 mmHg
Mean NBP:
20 - 150 mmHg
Diastolic NBP:
10 - 130 mmHg
Measuring Range
(Neonatal - 140mmHg)
Heart rate:
30 - 240 bpm
Systolic NBP:
30 - 130 mmHg
Mean NBP:
20 - 110 mmHg
Diastolic NBP:
10 - 100 mmHg
Default Inflation Pressure
Adult (270):
160 mmHg ±10 mmHg
Pediatric (180):
120 mmHg ±10 mmHg
Neonatal (140):
110 mmHg ±10 mmHg
Inflation Pressure After a Valid
Measurement
Adult (270):
Previous SYS + 25 mmHg (±10)
Pediatric (180):
Previous SYS + 25 mmHg (±10)
Neonatal (140):
Previous SYS + 30 mmHg (±10)
Inflation Pressure
After a Technical Alarm
Adult (270):
160 mmHg ±10 mmHg
Pediatric (180):
120 mmHg ±10 mmHg
Neonatal (140):
110 mmHg ±10 mmHg
Maximum Inflation
Pressure
Adult (270):
265 mmHg ±5 mmHg
Pediatric (180):
180 mmHg ±10 mmHg
Neonatal (140):
142 mmHg ±10 mmHg
Minimum Inflation
Pressure
Adult (270):
110 mmHg ±10 mmHg
Pediatric (180):
90 mmHg ±10 mmHg
Neonatal (140):
70 mmHg ±10 mmHg
Maximum Measurement Time
Adult (270):
2 min
Pediatric (180):
2 min
Neonatal (140):
90 sec
(60s French homologation)
Maximum Measurement Time
Including a Retry
Adult (270):
3 min
Pediatric (180):
--- (n/a)
Neonatal (140):
--- (n/a)
Software Safety Cut-Off
Adult (270):
273 ±3 mmHg
Pediatric (180):
215 ±3 mmHg (SWi)*
185-212 (SWh)*
Neonatal 140):
150 ±3 mmHg
Hardware Safety Cut-Off
Adult (270):
300 ±30 mmHg
Pediatric (180):
300 ±30 mmHg
Neonatal (140):
158 ±2 mmHg
Static Cuff Accuracy
±3 mmHg
Calibration Range
Adult and Pediatric:
10 - 260 mmHg ±3 mmHg
Neonatal:
10 - 150 mmHg ±3 mmHg
Non-Invasive Blood Pressure (NBP)
(continued)
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...