ECG
AND
HR S
AFETY
C
ONSIDERATIONS
VF6
Infinity Vista
8-17
To minimize interference from ESUs, we recommend the following:
z
Place the electrodes as far from the surgical incision as possible while
maintaining a clinically useful configuration.
z
Place the cable and lead wires as far from the ESU as possible and
perpendicular to the ESU cables.
z
Use an ESU neutral electrode with the largest possible contact area.
z
When possible, place the ESU neutral electrode close to and directly under the
surgical site, avoiding bony protuberances.
z
Replace the electrodes at regular intervals.
z
Read the instructions provided with the ESU for additional information.
z
For patients without a pacemaker, turn the pacer detection off. If pacer
detection is on, the ESU interference may be detected as pacer spikes that
display on the ECG.
The ECG waveform(s) and the HR value may be affected during electrosurgery.
However, after discontinuing the use of electrosurgery equipment, the waveform(s)
and value are displayed as normal.
Infusion pumps
Use of an infusion pump may cause artifact in ECG signals. To determine if the pump
is the source of electrical interference in the signal, turn it off, if possible. If the
artifact disappears, it was probably caused by the pump.
To minimize artifact and improve the signal, try the following:
z
Choose ECG lead(s) with the best signal for monitoring or replace the
electrodes.
z
Keep ECG cables away from the infusion pump and its wiring.
Defibrillators and Cardioversion
WA R N I N G :
z
B e f o r e a t t e m p t i n g a c a r d i o v e r s i o n , v e r i f y t h e t i m i n g o f
t h e s y n c p u l s e o n y o u r m o n i t o r.
z
N e v e r p l a c e t h e d e f i b r i l l a t o r pa d d l e s o v e r t h e E C G
e l e c t r o d e s o r c a b l e s . T h e d i s c h a r g e c a n b u r n t h e
pa t i e n t o r t h e c l i n i c i a n , o r f i b r i l l a t e t h e c l i n i c i a n a n d
n o t d e f i b r i l l a t e t h e pa t i e n t .
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...