C-16
Infinity Vista
VF6
C D
EFAULT
S
ETTINGS
AND
B
IOMEDICAL
S
UPPORT
NBP/SpO
2
Interlock
When NBP/SpO
2
Interlock is ON, SpO
2
alarms are disabled during an NBP
measurement in order to avoid nuisance SpO
2
alarms caused by impeded circulation
in the limb.
Turn NBP/SpO
2
Interlock on as follows:
1. Press the
Menu
fixed key.
2. Click on
Monitor Setup
.
3. Click on
Biomed
.
4. Enter the Biomed password and click the knob again.
5. Click on
Monitor Setup
.
6. Click on
NBP/SpO
2
Interlock
.
7. Turn the function
ON
or
OFF
.
ASY/VF Alarms
Set alarms for asystole and ventricular fibrillation as either
Always ON
or
Follow HR
.
If you select
Follow HR
, ASY and VF alarm settings follow those settings selected for
heart rate alarms in the Alarm Limits table. If you select
Always ON
, the monitor
always alarms for asystole and ventricular fibrillation, even if HR alarms and
arrhythmia monitoring are turned off.
Set ASY/VF Alarms as follows:
1. Press the
Menu
fixed key.
2. Click on
Monitor Setup
.
3. Click on
Biomed
.
4. Enter the Biomed password and click the knob again.
5. Click on
Monitor Setup
.
6. Click on
ASY/VF Alarms.
7. Select
Always ON
or
Follow HR
.
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...