D-12
Infinity Vista
VF6
D T
ECHNICAL
D
ATA
Interfering Substances
Carboxyhemoglobin may erroneously increase readings. The
level of increase is approximately equal to the amount of
carboxyhemoglobin present. Dyes, or any substance
containing dyes, that change usual arterial pigmentation may
cause erroneous readings.
Temperature
Input Connector
7-pin connector on MultiMed pod
Measurement Range
0
q
C to +50
q
C (32
q
F to 122
q
F)
Measurement Accuracy
±0.1 °C for a range of 0 °C to 50 °C
Probe Accuracy
±0.1 °C for a range of 0 °C to 50 °C
Total System Accuracy
±0.2 °C for a range of 0 °C to 50 °C
Average Response Time
< 2.5 seconds
Linearity
Linearity calculations to 0.02
q
C by the software algorithm.
Excitation Signal
DC source
Power to Probe
< 50
P
W (causing a potential self-heat error factor
of < 0.05
q
C)
Test Signals
– 5
q
C and 50
q
C ±0.1
q
C with nominal thermistor
Filter Characteristic
Low-pass filtering to minimize noise pickup.
Error Detection
> 50 °C: high temperature, possibly caused by a shorted
thermistor.
–3 °C: low temperature, possibly caused by an open
thermistor.
Self-test failure
Degree of Protection Against
Electric Shock
CF
Defibrillation Protection
In accordance with IEC 601-1A2
Non-Invasive Blood Pressure (NBP)
Parameter Display
Systolic, Diastolic, Mean
Measuring Method
Oscillometric technique
Modes of Operation
Manual (single measurement), Continuous (5 minutes), or
Interval
Pulse Oximetry (SpO
2
) with Masimo SET Pod
(continued)
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...