A
LARM
G
RADES
VF6
Infinity Vista
5-3
Serious Alarms
A serious alarm has the second highest priority. It alerts you to significant changes in
the patient’s condition other than life-threatening events as defined in the preceding
section. For example, if a monitored parameter falls below the selected alarm limits,
the monitor triggers a serious alarm. An unrecognized physiological condition, as
when a detected parameter is outside the monitor’s measuring range, triggers a serious
alarm also.
A serious alarm has the following characteristics:
z
The parameter box blinks and appears yellow.
z
The monitor produces two short tones continuously.
z
The monitor displays the alarm cause in both the parameter box and the
message area.
z
The monitor stops the alarm automatically when the condition ceases to exist.
z
The monitor stops the alarm when a new alarm of equal or higher grade is
triggered.
Advisory Alarms
An advisory alarms
has the lowest priority. It can result from one of three conditions:
z
When the monitor cannot produce a parameter value due to a technical
problem (e.g. a lead-off condition).
z
A patient cable or accessory failure (e.g. a blocked line in the pressure cuff).
z
The presence of persistent artifact.
An advisory alarm has the following characteristics:
z
The parameter box blinks and appears white.
z
The monitor produces a single tone followed by a 1½ second pause,
continuously.
z
The monitor displays the alarm cause in both the parameter box and the
message area.
z
The monitor stops the alarm automatically when the condition ceases to exist.
z
The monitor stops the alarm when a new alarm of equal or higher grade is
triggered.
NOTE:
Apnea events trigger a serious alarm in all monitoring modes.
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...