11-12
Infinity Vista
VF6
11 R
ESPIRATION
M
ONITORING
OxyCRG Monitoring (Neonatal Option)
The monitor offers a neonatal Oxycardiorespirogram (OCRG or OxyCRG) provided
the option is enabled on your monitor (for more information, see your Dräger
representative). Neonatal OxyCRG is a non-invasive procedure to display a beat-to-
beat heart rate (bbHR) trend, an SpO
2
trend, and a respiration waveform
simultaneously on a dynamically updated display. (OCRG monitoring requires that
sensors and electrodes for SpO
2
, HR, and respiration are attached to the patient.)
Displaying OCRG Waveforms
The OxyCRG waveforms are displayed in the second and third waveform channels.
The second waveform channel shows the beat-to-beat Heart Rate (bbHR) trend and
the SpO
2
trend waveforms. The bbHR trend has a range of 65 to 290. The SpO
2
trend
has a range from 70% to 100%. The third waveform channel shows a time-
compressed respiration waveform.
STEPS: Displaying the OCRG Waveforms
1. Make sure the Neonatal monitoring mode is selected.
2. Click on the
second
or
third waveform
channel.
3. Click on
Waveform
.
4. Select
OCRG
and click the knob.
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...