Overview
The monitor measures impedance respiration by passing a high-frequency current
between two ECG electrodes on the patient's chest. Electrical resistance (impedance)
between the electrodes varies with the chest’s expansion and contraction during
inspiration and expiration. The monitor can derive a respiration waveform and rate
from these impedance changes.
You can set the monitor’s breath detection sensitivity to Auto or Manual. If set to
Auto, the monitor automatically adjusts the breath detection sensitivity according to
the average size of the detected breaths. If set to Manual, the detection sensitivity is
based on the waveform amplitude you select (see the section
Respiration Monitoring
Display
, below).
The monitor is designed to:
z
Display the respiration waveform continuously.
z
Calculate the average respiration rate per minute.
z
Detect apnea in adult, pediatric, and neonatal patients.
Patient Preparation
Selecting and Preparing the Electrodes
Proper selection and preparation of the ECG electrodes is important for effective
respiration monitoring. See the chapter
ECG and Heart Rate
for information on
preparing and placing electrodes.
Preparing the Patient’s Skin
Follow the recommended instructions given in the chapter
ECG and Heart Rate
.
NOTE:
When using a 3-Lead ECG cable, you must display ECG lead II. If Lead I or
III is displayed instead, the monitor does not calculate a respiration rate and requests that
you select Lead II.
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...