13-8
Infinity Vista
VF6
13 N
ON
-I
NVASIVE
B
LOOD
P
RESSURE
As soon as you turn the interval mode on, the monitor starts an
NBP measurement. A bar graph appears in the NBP parameter box
to indicate the time left until the start of the next measurement.
In addition to measurements in the automatic cycle, you can take a
single measurement at any time. However, if the bar graph empties
before the end of your single NBP measurement, the next
automatic NBP measurement occurs only when the bar graph empties again. As a
result, the monitor skips one interval measurement.
If the monitor detects an error (while the NBP alarm is on), any attempts to take NBP
measurements (single or at intervals) generate an error tone. For more information,
see the chapter
Alarms and Messages
.
Inflation Mode
You must select the appropriate inflation limits for your patient. The following table
shows the monitor’s inflation values for each inflation mode:
Inflation
Mode
Initial
Inflation
(mmHg)
Inflation After a Valid
Measurement (mmHg)
Maximum
Inflation
(mmHg)
Minimum
Inflation
(mmHg)
Adult 270
160 ±10
Last Sys + 25 ±10
265 ±5
110 ±10
Pediatric 180 120 ±10
Last Sys + 25 ±10
180 ±10
90 ±10
Neo 140
110 ±10
Last Sys + 30 ±10
142 ±10
70 ±10
Note:
For patient category
Adult
, you can choose inflation modes
Adult 270
,
Pediatric 180
and
Neonatal 140
.
For patient category
Pediatric
, you can choose inflation modes
Pediatric 180
and
Neonatal 140
.
For patient category
Neonatal
, you can choose inflation mode
Neonatal 140
.
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...