13-3
NBP Safety Considerations
WA R N I N G :
z
C h a n g e s i n t h e c u f f p o s i t i o n r e l a t i v e t o t h e h e a r t l e v e l
a f f e c t m e a s u r e m e n ts . I f t h e c u f f i s n o t p l a c e d a t h e a r t
l e v e l , a d d + 1 . 4 m m H g f o r e a c h 2 c m a b o v e t h e h e a r t a n d
s u b t r a c t - 1 . 4 m m H g f o r e a c h 2 c m b e l o w t h e h e a r t . M a k e
s u r e t h a t t h e h o s e i s n o t k i n k e d o r o b s t r u c t e d . D o n o t
p l a c e t h e c u f f o n a l i m b w i t h a n i n f u s i o n l i n e .
z
I n s o m e c a s e s , r a p i d a n d p r o l o n g e d m e a s u r e m e n ts c a n
r e s u l t i n p e t e c h i a , i s c h e m i a , p u r p u r a o r n e u r o pa t h y.
D r ä g e r r e c o m m e n d s t h a t y o u a p p l y t h e c u f f
a p p r o p r i a t e l y a n d t h a t y o u c h e c k t h e c u f f s i t e r e g u l a r l y
w h e n m o n i t o r i n g a t f r e q u e n t i n t e r v a l s o r o v e r e x t e n d e d
p e r i o d s o f t i m e . I n a d d i t i o n , c h e c k t h e pa t i e n t f o r s i g n s
o f i m p e d e d b l o o d f l o w i n t h e l i m b .
z
N B P m e a s u r e m e n ts m a y n o t b e a c c u r a t e w i t h
c o n v u l s i v e pa t i e n ts o r pa t i e n ts w i t h t r e m o r s .
CAUTION:
Do not allow the hose or cuff to get in contact with fluids. Check the hose
and cuff frequently for signs of damage and debris. An obstruction in the hose may
cause the cuff to inflate and deflate improperly and may result in inaccurate readings.
NOTE:
To obtain accurate blood pressure readings, keep the limb and cuff
motionless. To protect patients from extremely high cuff pressures and extended cuff
inflation, the cuff deflates automatically in the following conditions:
z
The cuff pressure exceeds 273 mmHg in Adult 270 inflation mode, or 180
mmHg in Pediatric 180 inflation mode.
z
The cuff pressure exceeds 150 mmHg in Neo 140 inflation mode.
z
The measurement takes longer than 2 minutes for Adult 270 or Pediatric 180
inflation modes.
z
The measurement takes longer than 90 seconds (or longer than 60 seconds in
French NFC mode) for Neo 140 inflation mode.
z
A technical alarm has occurred.
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...