D-14
Infinity Vista
VF6
D T
ECHNICAL
D
ATA
Degree of Protection Against
Electric Shock
Type CF
Defibrillation Protection
per EN 60601-2-30 (IEC 601-2-30)
*SWi refers to instant measurements; SWh refers to measurements lasting more than 5 seconds and less than
15 seconds.
Invasive Blood Pressure (IBP1, IBP2)
Accessories
Transducers with a resistance of 300 to 2000
:
and an
equivalent pressure sensitivity of 5
P
V/V/mmHg ±10%
Input Connector
7-pin connector on monitor
Number of Channels
2 max (requires IBP options)
Sweep Speed
25 mm/s
Measured Parameters
Systolic, Mean and Diastolic blood pressure
Measurement Range
-50 mmHg to 399 mmHg
Measurement Accuracy
±2 mmHg or ±3% (whichever is greater) after successful zero
and calibration (exclusive of transducer).
Resolution
1 mmHg or 0.1 kPa
Response Time
d
35 s (to reach 90% of the change of the lowest pulse rate of
25 1/min.
Zero Balance Range
±190 mmHg
Zero Resolution
0.24 mmHg
Zero Accuracy
±0.48 mmHg
Bandwidth
DC to 16 Hz
Defibrillation Protection
In accordance with IEC 601-2-34
Degree of Protection Against
Electric Shock
CF
Non-Invasive Blood Pressure (NBP)
(continued)
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...