Reducing EMI
To reduce possible problems caused by EMI we recommend the following:
z
Use only Dräger-approved accessories.
z
Ensure that other products used in areas where patient monitoring and/or life-
support is used comply to accepted emissions standards (EN55011).
z
Try to maximize the distance between electromedical devices.
z
Strictly limit exposure and access to portable radio-frequency sources (e.g.,
cellular phones and radio transmitters). Be aware that portable phones may
periodically transmit even when in standby mode.
z
Maintain good cable management. Do not route cables over electrical
equipment. Do not intertwine cables.
z
Ensure all electrical maintenance is performed by qualified personnel.
Site of Operation
CAUTION:
The site of operation for the monitor must meet temperature, humidity, and
air pressure requirements. For details, see the product description in Appendix D.
WA R N I N G : D o n o t o p e r a t e t h e m o n i t o r i n p r e s e n c e o f
f l a m m a b l e a n e s t h e t i c m i x t u r e s w i t h a i r, o x y g e n , o r n i t r o u s
o x i d e . I f t h e d e v i c e s a r e u s e d w h e r e f l a m m a b l e a n e s t h e t i c
s u b s ta n c e s a r e u s e d , t h e p o s s i b i l i t y o f a n e x p l o s i o n c a n n o t
b e e x c l u d e d .
WA R N I N G : D o n o t u s e t h e m o n i t o r n e a r d e v i c e s w i t h
m i c r o w a v e o r o t h e r h i g h f r e q u e n c y e m i s s i o n s t h a t m a y
i n t e r f e r e w i t h t h e m o n i t o r ’s o p e r a t i o n .
WA R N I N G : I f f l u i d s a r e a c c i d e n ta l l y s p i l l e d o n t h e m o n i t o r, i t
s h o u l d b e r e m o v e d f r o m s e r v i c e i m m e d i a t e l y a n d t h o r o u g h l y
i n s p e c t e d b y y o u r B i o m e d t o e n s u r e t h a t t h e r e i s n o
c o m p r o m i s e i n e l e c t r i c a l s a f e t y.
CAUTION:
Place the monitor on a flat and stable surface to prevent it from falling. Do
not place the monitor into a cabinet, wall recess or similar enclosure during operation.
These units are convection cooled (no fan) and need adequate airflow to dissipate heat.
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...