Message: NBP
blocked Line
Parameter value: ***
There is a significant increase in
cuff pressure during the inflation
cycle.
Blocked output condition detected.
Cuff or tubing may be obstructed.
• Substitute cuff or tubing if
necessary.
Message: NBP
Overpressure
Parameter value: ***
The cuff pressure exceeds the
overpressure threshold.
• Call your Biomed.
Message: NBP Sys
<LL <value>
Message: NBP Sys
>UL <value>
Message: NBP Dia
<LL <value>
Message: NBP Dia
>UL <value>
Message: NBP Mean
>UL <value>
Message: NBP Mean
<LL <value>
Parameter value:
<value>
The patient’s pressure falls outside
the current alarm limits or these
limits are inappropriate for this
patient.
• Observe the patient.
• Change the current alarm limits
for this patient.
Message: NBP Out of
Range
Parameter value: ***
NBP values are above/below the
NBP measuring range.
• Check patient and treat, if
necessary.
• Verify that the inflation mode is
appropriate for the patient.
Change if necessary.
Message: NBP Low
Inflation Limit
Parameter value: ***
The patient’s systolic pressure is
above the current inflation limit.
• Change to a higher inflation limit.
Message: NBP Check
Cuff Size
Parameter value: ***
The wrong size NBP cuff is being
used on the patient.
• Verify the cuff size and placement
are proper for the patient.
• Verify all connections.
• Change the cuff if necessary.
Message: NBP Check
Hose Connection
Parameter value: ***
There is no significant increase in
cuff pressure during the inflation
cycle.
• Verify all connections.
• Change the hose if necessary.
Non-Invasive Blood Pressure (NBP)
(continued)
Message
Possible Cause
Suggested Action
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...