ECG and HR Safety Considerations
HR Alarm Settings
Set HR alarm limits on the Alarm Limits table (see the chapter
Alarms and Messages
).
If you turn the HR alarm off, the monitor displays a crossed bell icon in the HR
parameter box and the banner
HR Limits Off
above the first waveform channel. If
arrhythmia monitoring and alarms for asystole and ventricular fibrillation are also
disabled, the monitor displays the banner
HR, ASY, VF, Off
.
Neonatal ECG Monitoring
To obtain the most accurate heart rate count and waveform display, select the ECG
lead(s) with the smallest T wave and the largest monophasic QRS.
ECG 50/60 Hz Notch Filter Setting
When the line frequency is set incorrectly, the resulting 50 or 60 Hz noise on the ECG
signal can make the ECG waveform unreadable along with masking QRS complexes.
As a result, detection of the QRS complexes becomes unreliable. When these cannot
be sensed reliably, incorrect HR calculations may result, leading to both false positive
and false negative HR alarms.
Muscle Stimulators
NOTE:
z
If arrhythmia detection is on, alarms for asystole (ASY) and ventricular
fibrillation (VF) are automatically on and cannot be turned off.
z
By default, alarms for asystole and ventricular fibrillation are always on;
however, these alarms can be disabled in the password-protected Biomed menu
(see the appendix
Default Settings and Biomedical Support
).
WA R N I N G : I f t h e m o n i t o r d i s p l a y s t h e b a n n e r
H R , A S Y, V F
O ff
, t h e r e i s n o a l a r m a n n u n c i a t i o n f o r a s y s t o l e a n d
v e n t r i c u l a r f i b r i l l a t i o n .
WA R N I N G : N e o n a ta l Q R S c o m p l e x e s t h a t a r e b i p h a s i c a n d
l e s s t h a n 4 0 m s w i d e , o r m o n o p h a s i c a n d l e s s t h a n 2 0 m s
w i d e , a r e d i s p l a y e d s m a l l e r t h a n t h e i r a c t u a l a m p l i t u d e .
WA R N I N G : T h e u s e o f e l e c t r i c a l m u s c l e s t i m u l a t o r s m a y
i n t e r f e r e w i t h E C G a n d r e s p i r a t i on m o n i t o r i n g . I f t h i s o c c u r s ,
d i s c o n t i n u e t h e i r u s e .
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...