S
TORED
R
ECORDINGS
VF6
Infinity Vista
7-13
Stored Recordings
The monitor can store up to 10 alarm or timed recordings consisting of 20 seconds of
data each (about 10 seconds of pre-event/request data and about 10 seconds of post-
event/request data.) Recordings are stored when:
z
No recorder/laser printer is available or if the recorder/laser printer is
temporarily out of order (i.e. no recording paper).
z
You turned on the automatic storage of alarm recordings on the Alarm Limits
table (see the chapter
Alarms and Messages
).
When the recording storage is full, the monitor deletes the oldest unsaved recordings
as new recordings come in. You can view, print, save, and delete stored recordings via
the Event Recall screen (see below).
Upon successfully storing a recording, the monitor emits 2 beeps. To cancel a
recording storage in progress, press the
Record
fixed key on the monitor within 5
seconds of the start of the recording storage.
Event Recall
You can view 4 seconds of waveform data for each stored recording on the Event
Recall screen (4 seconds of pre-event/request data). The screen also shows the
parameter values at the time of the recording storage.
To call up the Event Recall screen:
1. Press the
Menu
fixed key.
2. Click on
Review
.
3. Click on
Event Recall
.
NOTE:
The monitor does not store continuous, trend, OCRG or diagnostic log
recordings. Continuous recordings are converted to timed recordings and then stored.
NOTE:
z
You can also access the Event Recall screen via the Fast Access fixed key or by
clicking on Monitor Setup > Recordings > Review in the Main Menu.
z
The Event Recall screen shows only 4 seconds of data for each stored
recording, although the monitor has actually stored 20 seconds of data for each
stored recording. If you print the stored recording, the recording strip contains
all 20 seconds of stored data.
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...