5-20
Infinity Vista
VF6
5 A
LARMS
AND
M
ESSAGES
Message: MultiMed
Disconnected
SpO2 and PLS
parameter value:
<blank>
The M
ULTIMED
/N
EOMED
pod is
disconnected from the monitor.
• Verify the pod connections.
Message: SpO
2
Light
blocked
Message: PLS Light
blocked
Parameter value: ***
The sensor does not detect the
light.
The sensor cannot detect a signal
that can be measured.
• Check the sensor.
• Make sure the light indicator is
not blocked.
• Remove any trace of nail polish.
• Move sensor location.
• Use different sensor.
Message: SpO
2
Weak Signal
Message: PLS Weak
Signal
Parameter value: ***
The sensor may be faulty.
The sensor cannot detect a signal
that can be measured.
• Remove bandages or rings that
may obstruct the signal.
• Reapply the sensor.
• Observe the waveform on the
screen. If you suspect a faulty
sensor, replace it.
Message: SpO
2
Motion
Parameter value: ***
Message: PLS Motion
Parameter value: ***
The monitor has detected motion
artifacts in the signal. The cause
can be one of the following:
The patient is moving
The patient is coughing
Hemodynamic interference
Hypothermia
Vascular compromises to the
extremity
• Move sensor to another location.
• Observe the resulting waveform
on the screen.
Message: SpO
2
No
measurement
Message: PLS No
measurement
Parameter value: ***
The monitor has been unable to
measure the parameter for the last
30 seconds.
• Calm the patient.
• Check the sensor placement.
• Move the sensor to a location with
less movement.
• If message persists, call
DrägerService.
Message: SpO
2
Regulate Error
Message: PLS
Regulate Error
Parameter value: ***
The monitor detects an excessive
amount of ambient light giving low
amplitude and inaccurate readings
as a result. The possible sources of
excessive light are:
Surgical or bilirubin lamp
Fluorescent lights
Infrared heating lamps
Sunlight
• Cover the sensor with an opaque
material such as a towel.
Pulse Oximetry (SpO2)
(continued)
Message
Possible Cause
Suggested Action
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...