5-12
Infinity Vista
VF6
5 A
LARMS
AND
M
ESSAGES
Central Alarms
If the bedside monitor appears on the central station’s display, the central station
announces all bedside alarms. The alarm grade (life-threatening, serious, advisory) is
determined by the bedside monitor’s setting. If the central station fails to indicate a
bedside alarm within 10 seconds, the bedside monitor reports a network alarm error
and alarms sound at their highest volume at the bedside monitor. Once the network
error is corrected, the bedside monitor’s master volume returns to the last user setting.
From the central station, you can remotely:
z
Set alarm limits.
z
Turn parameter alarms On or Off.
z
Silence active alarms.
z
Turn alarm recordings On or Off.
Changes made to the alarm limits at the central station take effect immediately at the
bedside monitor. For more information, see the Instructions for Use of the central
station.
NOTE:
z
When the monitor is connected to the Infinity network, you can turn the master
speaker volume off at the bedside monitor. If the monitor is operating as a
stand-alone device, you cannot turn the speaker volume off at the bedside
monitor (also see the
Monitor Setup
chapter).
z
During a network alarm error, the master speaker volume at the bedside monitor
can be lowered via the Monitor Setup menu. The network error message
remains on the screen until the error condition is resolved.
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...