M
ESSAGES
VF6
Infinity Vista
5-15
Message: ECG Artifact
HR parameter value:
***
ST<lead1>/ST<lead2>
parameter value:
<blank>
Patient’s movement.
Shivering, tremors.
Excessive signal noise.
Bad contact of electrodes.
Interference from auxiliary
equipment.
• Calm the patient.
• Check the electrodes.
• Apply the electrodes carefully.
• Secure the electrodes.
• Observe good skin preparation
techniques.
• Isolate the patient from auxiliary
equipment, if possible.
• If the artifact message is frequent,
contact your Biomed to check the
setting of the notch 50/60 Hz filter.
Message: HR Too
High
Parameter value: +++
The patient’s heart rate falls
outside the upper range (300 beats
per minute).
• Observe the patient carefully.
• Apply treatment if clinically
indicated.
Message: HR>UL
<value>
Message: HR <LL
<value>
Parameter value:
<value>
The patient’s heart rate falls
outside the current upper or lower
alarm limits.
The current alarm limits are
inappropriate for this patient
• Observe the patient carefully.
• Apply treatment if clinically
indicated.
• Change the alarm limits as
described in the “HR Alarms”
section.
Message: MultiMed
Disconnected
PVC parameter value:
<blank>
HR parameter value:
***
ST<lead1>/ST<lead2>
parameter value:
<blank>
The M
ULTIMED
/N
EOMED
pod is
disconnected from the monitor.
• Verify the pod connections.
Message: Pacer Off
Pacer detection is not selected.
• If you are monitoring a paced
patient, select the pacer detection
function.
Message: V Fib
Parameter value: VF
Detection of a ventricular
fibrillation. The ventricular rhythm is
chaotic.
• Observe the patient and treat if
clinically indicated.
• View each lead on the monitor
and check the amplitude. To
detect the QRS, the amplitude
should be at least 0.5 mV on the
mV scale.
• Reposition or change the
electrodes if the amplitude is still
low.
• Use good skin preparation.
Heart Rate, Arrhythmia, ST Segment Analysis
(continued)
Message
Possible Cause
Suggested Action
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...