A
SSIGNING
A
LARM
G
ROUPS
VF6
Infinity Vista
5-11
Assigning Alarm Groups
If a monitor is part of the Infinity network, alarm messages from other networked
monitors appear in the message area at the bottom of the screen (i.e.
BED 20: ECG
Leads Off
). In order to limit the number of messages from remote bedside monitor,
you can group bedside monitors into separate alarm groups so that only messages
from monitors within the same alarm group are shared. If you do not want to display
any remote alarm message at a particular bedside monitor, place that monitor in its
own alarm group.
STEPS: Assigning Alarm Groups
1. Press the
Menu
fixed key.
2. Click on
Monitor Setup
.
3. Click on
Alarm Groups
.
4. Click the knob, dial in the number of the desired alarm group, and click the
knob again.
CAUTION:
The user must confirm at each individual bedside that the monitor is
assigned to the desired alarm group.
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...