D-8
Infinity Vista
VF6
D T
ECHNICAL
D
ATA
Pacer Pulse Detection Level
Amplitude:
± 5 to ± 700 mV
Width:
0.1 to 2.0 ms
Separation between pulses:
t
30 ms
Recharge time constant:
4 to 100 ms
Over/Undershoot:
0.025 a
p
2 mV maximum
Lead Off Sensing Current
<50 nA DC from LL, LA, RA and V Leads
Arrhythmia
Measurement Range
0 to 300 beats/minute
Display Resolution
1 beat/minute
Accuracy
± 5 beats/minute or ± 10% of the rate, whichever is greater
Response Time
< 4 seconds
ST Segment Analysis
Measurement Range
-15 mm to +15 mm (-1.5
mV to +1.5 mV)
Accuracy
±1mm/0.1mV RTI
Isoelectric Point Range
From start of the averaged ST complex to Fiducial Point, in
increments of 4 ms
ST Point Range
From Fiducial Point to end of the averaged ST complex, in
increments of 4 ms
Display Resolution
0.1 mm/0.01mV (10
P
V)
Response Time
15 seconds
Respiration
Method
Impedance Pneumography
Sensing Electrodes
RA and LL (Lead II)
Number of Channels
1
Sweep Speed
6.25 mm/s
ECG
(continued)
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...