5. With the rotary knob, move the position of the ST measuring point to the
desired location and click the knob. As you dial, the ST deviation values to
the right of the ST waveforms change according to the new measuring point
location.
ST Reference Complex
The averaged ST segment waveform(s) displayed in the ST menu can be saved as
reference complexes. The reference waveforms can then be superimposed on the
current ST segment waveform to highlight changes in the ST segment since the last
save. Upon saving, the reference complex is time-stamped and available for display.
STEPS: Displaying the ST Reference Complex
1. Click on the
ST
parameter box.
2. Click on
Ref
to turn the display of the reference complex ON or OFF
.
STEPS: Saving the ST Reference Complex
1. Click on the
ST
parameter box.
2. Click on
Save
. Date and time of saving appear next to the Save option.
NOTE:
If you change an ECG lead and access the ST Setup menu, you lose the saved
ST reference complex. The saved ST reference complex is retained if you change a lead
and later reselect it without entering the ST Setup menu.
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...