
O
VERVIEW
VF6
Infinity Vista
D-3
Basic System Components
Monitor Base Unit
Dimensions
196 mm height (7.7 in)
267 mm width (10.5 in)
147 mm depth (5.8 in)
Weight
3.87 kg (8.54 lb) with lead acid battery
3.67 kg (8.10 lb) with lithium ion battery
3.32 kg (7.32 lb) without battery
Materials
Plastics: ABS, FR 110
Printed Circuits: glass/epoxy, lead/tin solder, copper etch,
lithium battery
Battery: sealed lead acid, or lithium ion (option)
Heatsink: cast aluminum
NBP assembly: silicon tubing, steel, copper wire
Packing: corrugated cardboard, urethane foam
Disposal
All materials must be disposed of or recycled properly and in
accordance with local regulations. There are no known
special disposal requirements for any accessories.
Battery Pack
Lead Acid: PANASONIC LC-T121R8PU or equivalent
Li-ion:
Dräger Li+ Battery Pack
Protection Class
Internally powered, for use with specified Class I power
supply
Mode of Operation
Continuous (with AC adapter)
DC Input
11 – 15 VDC, 2.5A
Max. Power Consumption
52 Watts
Battery Operating Time
(with ECG, SpO2, Respiration
waveforms and NBP in a 15-
minute interval mode starting with
a fully charged battery)
Lead acid:
75 minutes
Lithium ion:
210 minutes in Bright Screen Mode and 240
minutes in Dim Screen Mode
Battery Recharging Time
Lead acid:
5½ hours, typical.
Lithium ion:
8 hours, typical.
Patient Leakage Current
< 10
P
A @ 110 V and 60 Hz
< 10
P
A @ 220 V and 50 Hz
QRS Sync. Output
Output will go high for 100 ms every time a QRS is detected.
QRS detected: +12 V ±5%, 5.1 K
:
source impedance.
Output low (no QRS): < .8 V @ 30 mA sync current.
Alarm Output
12 V 560
:
output for external alarm indicator (Nurse Call
System)
Recorder
UART output to interface with an R 50 Series recorder
through the interface plate.
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...