Zeroing and Calibration Check
The
Zero
field in the IBP setup menu:
z
Displays the date and time of the last zeroing procedure.
z
Allows you to zero the pressure transducer before entering a calibration
factor.
Pressure transducers are sensitive and their calibration may change significantly
following a mechanical shock or an overpressure in the system (e.g., after drawing
blood or injecting drugs).
If you are using reusable transducers, it is necessary to calibrate them with a mercury
manometer to determine the calibration factor. This calibration procedure is typically
the responsibility of your Biomed.
The calibration check procedure described in this section consists of adjusting the
monitor by entering the calibration factor given by your Biomed, if you are using
reusable transducers.
If you are using disposable transducers, use 100 as the calibration factor. Disposable
transducers are all pre-calibrated.
See the table on
Zero and Calibration Check Troubleshooting
(below) to help you
during the procedure.
STEPS: Zeroing and Entering a Cal. Factor
1. Click on the IBP1 parameter box.
2. Re-align the transducer to the patient’s heart level.
NOTE:
Before zeroing, make sure the transducer is at heart level. It is necessary to
zero the transducer immediately after the introduction of the catheter in the patient’s
vascular system and before monitoring. You should also zero the transducer once a day
or after changing the tubing or the dome of the transducer.
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...